3 Device-specific Menus Volume 6
Select the setting for RADIATION INTERFERENCE.
Click
TRANSFER to upload the setting to the Uni-Probe.
PULSE RATE LIMITS Tab
Parameters Input Description
Interference a No Detection: Turns the radiation interference detection off.
On: Turns the radiation interference detection on.
Parameters Input Description
Sensor value max a Upper pulse rate limit
Enter “
0” to disable this function. This function is enabled only for special
applications. Enter a value unequal to zero to enable the upper threshold.
If the current pulse rate exceeds the upper threshold, the measurement is
halted until the pulse rate has dropped below this threshold again. If the
measurement is stopped, the measured value display and the current
output are held. The count rate where switching takes place is the current
non-averaged count rate.
Sensor
value max
Sensor
value min
Measurement
halted
Measurement
halted
CPS
t