Volume 7 3 Parameter Overview
7
LE_DIGITAL_INPUT_ENTRY Selection field for digital input.
The input can be set to Hold or Empty. If you select Hold the measured
value is held as long as the digital input is closed. If it is set to Empty, an
empty adjustment is carried out if the digital input is closed for a short
time.
LE_DIGITAL_INPUT_FUNCTION Enables or disables the digital input.
LE_DIGITAL_INPUT_TEST_VALUES
LE_DI1_STATE
LE_DI2_STATE
LE_DI3_STATE
Record. Shows the current status of the digital inputs.
Status input 1
Status input 2
not used
LE_DIGITAL_OUTPUT_ENTRY Selection field for one of the following digital outputs: DO 2, 3 or 4. Allows
the parameter setting for the selected relay in the record
LE_DIGITAL_OUTPUT_SETUP. See section 5.7 on page 7–697.
LE_DIGITAL_OUTPUT_SETUP
LE_FUNCTION
LE_THRESHOLD
LE_HYSTERESIS
Record. Defines the values for the DO 2 to 4. DO 1 cannot be changed
(error relay). See section 5.7 on page 7–697.
Selection of the relay function
Threshold value for triggering the relay function
Hysteresis for triggering the relay function
LE_DIGITAL_OUTPUT_TEST_MODE Enables or disables the test mode for the digital output.
LE_DIGITAL_OUTPUT_TEST_VALUES
LE_DO1_STATE
LE_DO2_STATE
LE_DO3_STATE
LE_DO4_STATE
Record. Allows you to enter the test value Open or Closed to switch DO 1
to 4. LE_DIGITAL_OUTPUT_TEST_MODE must be enabled to switch the
DO.
Status DO 1
Status DO 2
Status DO 3
Status DO 4
LE_DO_TEST_MODE_STATUS Shows if the test mode for the DO is enabled or not. It is enabled via
LE_DIGITAL_OUTPUT_TEST_MODE.
LE_ERROR_HANDLING Defines the behavior if an error occurs. See section 7.1 on page 7–715.
LE_HOLD_MEASUREMENT The current process value is held. This function is relevant only for servic-
ing.
LE_INTERFERENCE_DETECTION
LE_DETECTION_MODE
LE_DETECTION_THRESHOLD
LE_DETECTION_HOLD_TIME
Record. Allows the detection and suppression of short-term interferences
caused by interfering radiation sources. See section 5.5 on page 7–696.
Enables or disables the detection
Threshold value for starting the suppression of measured values
Time period for which the measurement is held after radiation interfer-
ence detection.
LE_ISOTOPE Type of isotope used
LE_LEVEL_OFFSET Function not implemented
LE_LEVEL_UNIT Unit for the process value LE_LEVEL. This information is identical with
LE_PRIMARY_VALUE_UNIT. Modifying one value will also change the
other.
Parameter name Description