3 Parameter Overview Volume 7
LE_LEVEL
VALUE
Status
Level value
Process value in the unit defined in LE_LEVEL_UNIT. The value is identical
with VALUE of PRIMARY_VALUE. This value is passed on to the analog
output block, if AI VALUE: Level has been selected there.
Status information regarding the process value.
LE_LIN_DIAMETER Measuring path in the vessel. Used only for a one-point calibration.
Regarding one-point calibration, see section 4.4 on page 7–682, regard-
ing measuring path, see section 6.3 on page 7–705
LE_LIN_TYPE Defines the type of linearization: Linear or Exponential. See section 6.3
on page 7–705.
LE_MAX_SENSOR_VALUE If you are using a value greater than zero, the measured value is held as
long as the value indicated is exceeded by the unaveraged pulse rate
(LE_SENSOR_LIVE_VALUE). The measurement will not switch to the
alarm status. See section 5.8 on page 7–698.
LE_MIN_SENSOR_VALUE If you are using a value greater than zero, the measured value is held as
long as the value indicated is below the unaveraged pulse rate
(LE_SENSOR_LIVE_VALUE). The measurement will not switch to the
alarm status. See section 5.8 on page 7–698.
LE_NUMBER_OF_CAL_POINTS Number of calibration points. This value has to be set before entering the
calibration table. The value can also be changed later if a new calibration
is carried out. See section 4.5 on page 7–684.
LE_NUMBER_OF_DETECTORS Number of detectors used in the multi-detector mode. See section 5.2 on
page 7–691.
LE_OPERATING_MODE Current operating mode of the Uni-Probe. See section 7.3 on page 7–716.
LE_PLATEAU_DATA_ENTRY Selection of the value pair of the plateau measurement displayed in
LE_PLATEAU_DATA. See section 5.1 on page 7–689.
LE_PLATEAU_DATA_NUMBER Number of value pairs obtained in the course of a plateau measurement.
The value is a result of the settings in LE_PLATEAU_SETUP. See section
5.1 on page 7–689.
LE_PLATEAU_DATA
LE_PLATEAU_DATA_HIGH_VOLTAGE
LE_PLATEAU_DATA_SENSOR_VALUE
Record. Shows the value pair selected with LE_PLATEAU_DATA_ENTRY.
See section 5.1 on page 7–689.
HV value
Pulse rate at this HV value
LE_PLATEAU_HIGH_VOLTAGE Actual HV value. See section 5.1 on page 7–689.
LE_PLATEAU_PULSE_RATE Actual pulse rate during the plateau measurement. See section 5.1 on
page 7–689.
LE_PLATEAU_SETUP
LE_START_HV
LE_END_HV
LE_HV_STEP
LE_MEAS_TIME
Record. Used for the setting of the plateau measurement. See section 5.1
on page 7–689.
HV start value
HV end value
HV step width
Duration of the measurement for each step of the plateau measurement
LE_PLATEAU_STATUS Shows Read In On if a plateau measurement is running. See section 5.1
on page 7–689.
LE_PRIMARY_VALUE_UNIT Unit for the process value PRIMARY_VALUE. The information is identical
with LE_LEVEL_UNIT. Modifying one value will also change the other.