Web Interface 8
8.3.1.2 Additional information about the error message code
The error message code indicates which spectrometer component is defective or mal-
functioning. The error message codes are grouped as follows:
8.3.1.3 Additional information about the importance level of error/warning messages
The error/warning messages are divided into static and non-static messages.
Static error/warning messages:
Error message code Spectrometer component
10000 - 10999 Scanner (interferometer block)
20000 - 20999 Detector
30000 - 30999 Laser
40000 - 40999 Source
50000 - 50999 Automation units
60000 - 60999 Instrument
Table 8.4: Error message code
Importance level Description
S (Static fatal error, non-
resettable)
R (Resettable static fatal
error)
Messages of the importance levels S and R indicate a
static fatal error which prevents a measurement from
being executed, or cancels a running measurement.
In this case, the OPUS status light is red and an error
message is displayed in OPUS.
The error remains until the problem has been solved or
the unit causing the error is no longer selected (e.g.
instead of a detector which is not cooled down to oper
-
ating temperature, another detector has been selected).
Table 8.5: Static error/warning messages