148
IFS 125M User Manual Bruker Optik GmbH
Web Interface 8
In case of static error/warning messages, there are resettable and non-resettable mes-
sages. If a resettable error/warning message is displayed, send the corresponding com-
mand to remove the error message.
In case of a non-resettable error/warning message, first eliminate the cause of the error, 
then send the corresponding command.
Example:
To remove the message shown in figure 8.6, either send a direct command to bring the 
motorized unit in its default position or connect the respective motor first and then send 
the direct command CHN=1 once again.
V (Static warning, non-
resettable)
U (Resettable static 
warning)
Messages of the importance levels V and U indicate a 
static warning which does not prevent, or cancels a 
measurement.
In this case, the OPUS status light is yellow, and an 
error message is displayed in OPUS. 
The error state remains until the cause of the problem 
has been eliminated. 
Importance level Description
Table 8.5: Static error/warning messages
Figure 8.6: Example of error/warning message