Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
A-2(2)
Document no. M1125636-05
General
22. Communication lines
+Reset RS485 (pin 1))
>10 k
-Data RS485 (pin 5))
>10 k
+Data RS485 (pin 6)
>10 k
-Reset RS485 (pin 8)
>10 k
OK N.A. Fail OK N.A. Fail
23. Fan
24. Starting
25. Module
communication
26. Real time clock
27. Loudspeaker sound
28. ECG out
29. Pressure out
30. Monitor software L-
31. Content of service log
32. Voltages
+15 V 14.20...15.60 V
-15 V -14.00...-15.50 V
+15 VD) 14.10...15.60 V
+5 V 4.70...5.40 V
33. Watchdog circuitry
34. Trend retaining
35. Service reset button
36. Recovering from
power loss
37. Electrical safety check
38. Functioning after
electrical safety check
39. Final cleaning
Notes
Used spare parts
Signature