18-6 Power Model 831 Manual
The default value is “Alkaline”.
WARNING:Do not mix Alkaline and NiMH batteries.
WARNING:Do not mix batteries from different
manufacturers
WARNING:Replace all four batteries when installing
fresh cells
WARNING:The correct battery type must be specified,
as described in "Battery Type" on page 18-5, based on
the battery type installed. Otherwise, serious damage,
injury or fire can occur when the battery type is set to
NiMH but Alkaline or Lithium batteries are installed
because the internal charger will be enabled. Alkaline or
Lithium batteries must not be charged.
Auto-Off Time
Auto-Off time is the duration of time the instrument will
stay on when no activity is occurring: button presses,
running a measurement, USB communications, etc.
Pressing the
0 (ON / OFF) key will return the instrument
and the display to the state it was in when the Auto-Off time
expired.
The auto-off feature is ignored when connected to external
power (assumed to mean when not on internal batteries
which includes USB and External Power). When the unit is
connected to USB power, the feature is ignored but when it
is connect to external power (12 Vdc) it is not ignored.
To set the Auto-off Time, highlight the Auto-Off Time data
field and press
5 to open the Auto-Off Time Menu, shown
in FIGURE 18-8.
FIGURE 18-8 Auto-Off Time Menu
Highlight the desired time and press
5 to make a selection.
The default value is “Never”.