29.17 Automation
■■■■■■■■■■■■■■■■■■■■■■
478
■■■■■■■■
900 Touch Control
Subsequence type
Selection Sample sequence | Start sequence | End
sequence | Stop sequence | Cond. sequence
Default value Sample sequence
Sample sequence
The subsequence is carried out with every determination.
Start sequence
This subsequence is only carried out at the start of a sample series
when the autostart counter = 1.
End sequence
This subsequence is only carried out at the last sample of a series when
the autostart counter has reached the setpoint value (number of autos-
tarts).
Stop sequence
The stop sequence is only carried out when canceling a method
because of one of the following cases: Manual stop with [ ], stop
because of an error, stop via remote signal.
Cond. sequence
This subsequence is carried out directly before the conditioning. This
gives you the possibility to start e.g. a connected Polytron
®
high-fre-
quency homogenizer before the conditioning as well as to set its stir-
ring rate. Use the command CONTROL RS for this.
Cycles
Number of directly consecutive repetitions of the subsequence.
Input range 0 to 999
Default value 1
Selection Calibration
Calibration
For automatic calibrations with a Sample Processor (see chapter 32.7,
page 497).
[Note]
Entering a note on the subsequence.
Dialog "Subsequence options / Note"
A short text can be entered in this dialog, e.g. for the description of the
subsequence commands or for the application purpose.
[Display options]
Definition when the note is being displayed.