29 Parameters
900 Touch Control
■■■■■■■■
417
Measured value (measuring mode T):
Input range –20.0 to 150.0 °C
Default value 150.0 °C
Measured value (measuring mode Cond):
Input range 0.0001 to 999.9999 mS/cm
Default value 999.9999 mS/cm
Time:
Input range 0.0 to 999,999.9 s
Default value 999,999.9 s
Volume:
Input range
0.00000 to 9,999.99 mL
Default value 9,999.99 mL
29.12.5 Rate evaluation (EVAL RATE)
The rate evaluation provides the median dosing rate in a defined range
(so-called window). The median dosing rate is determined by means of lin-
ear regression through a minimum of three measuring points. If the dos-
ing is carried out with only one dosing device and if more than one cylin-
der volume is dosed, then the median dosing rate will be reduced as a
result of the filling time.
When the command is opened for the first time, a window over the entire
time range is already defined. In contrary to the potentiometric evaluation
of a DET/MET titration the individual windows may overlap. A maximum
of 9 windows can be defined.
[New]
Define a new window.
[Delete]
Delete the selected window.
[Edit]
Edit the settings of the selected window.
Lower limit
Time for the lower limit.
Input range 0 to 999999 s
Default value 0 s
Upper limit
Time for the upper limit.