29 Parameters
900 Touch Control
■■■■■■■■
397
NOTICE
Deactivate the Stir solution during measurement parameter when
you carry out the calibration with an 814 USB Sample Processor or with
an 815 Robotic USB Sample Processor XL. The stirrer control for the
tower stirrer is ineffective with these instruments.
Stir solution during measurement
on | off (Default value: on)
If this parameter is activated, the stirrer will be switched on and off auto-
matically at the start or end of the measurement.
Stir before meas.
This parameter can only be edited when the parameter Stir solution
during measurement is deactivated.
If the stirrer is switched off during the measurement, it will be stirred for
the time entered before measuring.
Input range 0 to 999999 s
Default value 0 s
Pause before meas.
This parameter can only be edited when the parameter Stir solution
during measurement is deactivated.
If it is stirred before the measurement, a waiting time can be defined here
before starting the measurement. No stirring or measuring occurs during
this time.
Input range 0 to 999999 s
Default value 0 s
29.10.3 Calibration parameters (CAL Cond)
Under [Calibration parameters], the parameters influencing the run of
the entire calibration are defined.
Signal drift
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute.
Input range 00.001 to 999,000 (mS/cm)/min
Selection off
Default value off