29 Parameters
900 Touch Control
■■■■■■■■
311
Switch off automatically
on | off (Default value: on)
If this parameter is activated, the stirrer will be switched off automatically
at the end of the titration, measurement, etc.
29.3 Endpoint titrations (SET)
29.3.1 Start conditions
The parameters that are carried out before the start of titration are
defined under [Start conditions].
Pause 1
Waiting time, e.g. for the electrode to settle down before a start volume is
added.
Input range 0 to 999999 s
Default value 0 s
Start volume
Volume that is dosed prior to the start of the titration.
Input range 0.00000 to 9999.99 mL
Default value 0.00000 mL
Dosing rate
Rate at which the start volume is dosed.
Input range 0.01 to 166.00 mL/min
Selection maximum
Default value maximum
maximum
The maximum dosing rate depends on the cylinder volume (see chap-
ter 32, page 492).
Pause 2
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume.
Input range 0 to 999999 s
Default value 0 s
Initial measured value
The [Initial meas. value] is measured before carrying out the start con-
ditions. This value is used for the titration direction determination
when the titration direction is set to auto and only 1 EP is defined.