13 GLP manager
900 Touch Control
■■■■■■■■
139
13.5 System monitoring
13.5.1 Service interval
Main dialog: System GLP manager Monitoring Service inter-
val
In the dialog GLP manager / Service interval, you can define the time
interval for the maintenance of the system by the Metrohm Service
department. The service interval is checked each time the system is
started.
Last service
Date on which the last servicing was carried out.
Format: YYYY:MM:DD
Monitoring
on | off (Default value: off)
If this parameter is activated, then the time interval after which system
maintenance must be carried out again will be monitored.
Service interval
If you define a time interval for the system maintenance, then the date in
Next service will be tracked automatically.
Input range 1 to 999 days
Default value 999 days
Next service
If you specify a date for the next system maintenance, then the Service
interval will be tracked automatically.
Format: YYYY:MM:DD
13.5.2 Backup interval
Main dialog: System GLP manager Monitoring Backup inter-
val
In the dialog GLP manager / Backup interval, you can define the time
interval for backups. The backup interval is checked each time the system
is started.
Last backup
Date on which the last backup was created.
Format: YYYY:MM:DD