31.1 Editing methods
■■■■■■■■■■■■■■■■■■■■■■
484
■■■■■■■■
900 Touch Control
31 Troubleshooting
31.1 Editing methods
Problem Cause Remedy
The method com-
mand cannot be
inserted.
The method command is
disabled.
In the dialog Dialog options / Command
list, activate the desired command (see "Block-
ing unneeded commands and fixed keys",
page 25).
No control instrument
which supports the
method command is con-
figured in the device man-
ager.
Add a control instrument which supports the
desired method command to the list of devices
(see chapter 11, page 79).
The additional eval-
uations (EVAL com-
mands) cannot be
inserted.
A titration or measuring
command to which the
evaluation can be applied
is missing.
1. Insert a titration or measuring command.
2. Insert the EVAL command directly after it.
An EVAL command
is depicted in red in
the command list.
The associated titration or
measuring command has
been deleted or replaced
by a different one, e.g. DET
pH by DET U.
1. Delete the EVAL command.
2. Insert a titration or measuring command.
3. Insert the EVAL command directly after it.
31.2 Sample table
Problem
Cause Remedy
The method cannot
be entered in the
sample table.
The option Use sample
assignment table is acti-
vated. The method
assigned to the sample
identification will be loa-
ded automatically at the
start of the determination.
Deactivate the option (see chapter 15.1, page
147).
31.3 Results/Statistics
Problem
Cause Remedy
No statistics are car-
ried out for a result.
Statistics is not activated. In the method options, activate the option
Statistics (see chapter 16.5, page 172).