29.12 Evaluations (EVAL)
■■■■■■■■■■■■■■■■■■■■■■
410
■■■■■■■■
900 Touch Control
EVAL FIX-EP (fixed endpoints)
Measured quantities:
Measured value
Time
Volume
EVAL MIN/MAX (minimum/maximum)
EVAL RATE (average dosing rate)
Additional evaluations for measurements
The following additional evaluations are possible:
EVAL FIX-EP (fixed endpoints)
Measured quantities:
Measured value
Time
EVAL MIN/MAX (minimum/maximum)
EVAL BREAK (break points)
NOTICE
The commands EVAL are not available for concentration measurements
(MEAS Conc).
29.12.1 Fixed endpoint evaluation (EVAL FIX-EP)
For the fixed endpoint evaluation, the associated values are interpolated
from the measuring point list for one quantity (measured value, volume,
etc.). Up to nine fixed endpoints can be evaluated with one command.
Fixed quantity
Selection of the quantity to which the associated value is interpolated
from the measuring point list.
Selection Measured value | Time | Volume | Quantity
Default value Measured value
Quantity
Amount of iodine/bromine created (only for KFC/BRC).
Fixed EP1 at
The fixed endpoint must lie between the first and the final entry in the
measuring point list.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
instrument used.