29 Parameters
900 Touch Control
■■■■■■■■
285
29.1.3 Stop conditions
The conditions for canceling the titration are defined under [Stop condi-
tions].
Stop volume
The titration is stopped when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.
Input range 0.00000 to 9999.99 mL
Default value 100,000 mL
Selection off
Stop meas. value
The titration is canceled when the specified measured value has been
reached since the start of the titration.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
instrument used.
Measuring mode pH:
Input range –20.000 to 20.000
Selection off
Default value off
Measuring mode U, Ipol:
Input range
–2000.0 to 2000.0 mV
Selection off
Default value off
Measuring mode Upol:
Input range
–200.0 to 200.0 µA
Selection off
Default value off
Stop EP
The titration is stopped when the specified number of equivalence points
has been found.
Input range 1 to 9
Default value 9
Selection off