29.1 Dynamic equivalence point titrations (DET)
■■■■■■■■■■■■■■■■■■■■■■
286
■■■■■■■■
900 Touch Control
Volume after EP
This volume will be added when the number of equivalence points
defined under Stop EP has been found. The curve shape after the equiva-
lence point can also be seen this way.
Input range 0.01000 to 9999.99 mL
Selection off
Default value off
Stop time
The titration is stopped when the specified time has elapsed since the start
of the titration.
Input range 0 to 999999 s
Selection off
Default value off
Filling rate
Rate at which the dosing cylinder is filled.
Input range 0.01 to 166.00 mL/min
Selection maximum
Default value maximum
maximum
The maximum filling rate depends on the cylinder volume (see chapter
32, page 492).
29.1.4 Potentiometric evaluation
The parameters for the evaluation of the titration curve are defined under
[Potentiometr. evaluation].
EP criterion
The Equivalence point Recognition Criterion (ERC) found is compared with
this value. Equivalence points whose ERC is less than the value defined
here will be ignored.
Input range 0 to 200
Default value 5
EP recognition
This parameter allows you to filter out only the equivalence points that are
being sought.
Selection all | greatest | last | ascending | descending |
off
Default value all