29 Parameters
900 Touch Control
■■■■■■■■
361
Stop rate
The titration is stopped when the current titration rate is lower than the
value entered. Checking this parameter is not started until 10 s after the
start of the command.
Input range 0.01 to 166.00 µL/min
Selection off
Default value off
Filling rate
Rate at which the dosing cylinder is filled.
Input range 0.01 to 166.00 mL/min
Selection maximum
Default value maximum
maximum
The maximum filling rate depends on the cylinder volume (see chapter
32, page 492).
29.7.5 Monitoring
Under [Monitoring], the parameters for monitoring the following quanti-
ties are defined:
Measured value
Dosing rate (an average rate)
Temperature
Monitoring is not used for parameterization of the control behavior.
Remote signals or RS-232 commands can be assigned to the monitored
quantities (see "Communication", page 365). These remote signals or
RS-232 commands are sent if a limit has been infringed. This can be used,
for example, to switch a thermostat on/off if necessary.
The parameters are identical for all monitored quantities.
Monitoring
on | off (Default value: off)
Enabling and disabling the monitoring function. The monitoring of the
dosing rate is not started until 10 s after the start of the titration.
Lower limit
When the monitored quantity falls below this value, the action defined in
the following is triggered.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges