29.4 Volumetric Karl Fischer titrations (KFT)
■■■■■■■■■■■■■■■■■■■■■■
332
■■■■■■■■
900 Touch Control
Delay reconditioning
on | off (Default value: off)
If this parameter is activated, the conditioning will not start before all the
method commands have been processed. Otherwise, the conditioning will
immediately start after the titration command.
NOTICE
This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.
Example: Emptying the titration cell followed by adding new working
medium.
29.4.6 Cell
The monitoring of the reagent used is activated or deactivated under
[Cell].
Reagent monitoring
Selection of the reagent from the list of reagents. The selection depends
on the titration mode. Reagents are defined under System Reagents.
Selection Selection of configured reagents | off
Default value off
off
The reagent monitoring is not active.
29.4.7 Control instrument
The control instrument the determination is carried out with is selected
under [Control device]. Control instruments are defined under Sys-
tem Device manager.
NOTICE
This button is only displayed when several control instruments have
been configured.
Control device
Selection of the control instrument from the list of devices. Only those
devices are displayed which are able to carry out the command. The con-
trol instruments do not have to be connected. This allows to transfer
methods more easily from one system to another.