16 Methods
900 Touch Control
■■■■■■■■
175
Selection no | Selection of configured sample racks
Default value no
Increase sample variable automatically
on | off (Default value: on)
The sample variable describes the current position of the sample on the
rack of the Sample Processor. If this option is activated, then the sample
variable is automatically increased by 1 at the end of each determination.
This option must be deactivated if the sample variable is to be altered spe-
cifically with a SAMPLE command.
Request rack position at start
on | off (Default value: off)
If this option is activated, then the rack position of the first sample to be
processed is requested.
16.5.2 Stop options
Main dialog: Edit parameters ▶ Method options ▶ Start/Stop
options ▶ Stop options
In the dialog Method options / Stop options, you can define the
actions which are carried out when a method is canceled.
The method can be canceled as follows:
■ Manual stop with the fixed key [
]
■ Stop because of an error
■ Stop by remote signal via the Control Remote Box
Switch off pumps
on | off (Default value: on)
If this option is activated, then all of the connected pumps will be
switched off.
Switch off stirrers
on | off (Default value: on)
If this option is activated, then all of the connected stirrers will be
switched off.