31 Troubleshooting
900 Touch Control
■■■■■■■■
485
Problem Cause Remedy
No statistics variable has
been assigned to the cal-
culation.
In the result options (command CALC), define
a variable for the mean value (see "Variable for
mean value", page 153).
Statistics is not activated. In the control dialog, activate the Statistics
option (see chapter 18, page 186).
A result has been
removed from the
statistics, the mean
value has not been
reassigned to the
common variable or
to the TITER varia-
ble.
Assignment is not carried
out automatically with ret-
roactive modifications.
Recalculate the determination manually (see
chapter 24, page 214).
The result is not dis-
played in the result
table.
The column display in the
result table is incorrectly
configured.
Modify the settings for the display accordingly
(see chapter 26.1, page 237).
The parameter Save result
in result table is not acti-
vated.
In the result options (command CALC), acti-
vate the parameter (see "Save result in result
table", page 154).
31.4 Printing
Problem
Cause Remedy
Tapping on the fixed
key [ ] has no
effect.
The fixed key is disabled. In the Dialog options / Fixed keys dialog,
activate the Print option (see chapter 7.2,
page 24).
A determination is running. Wait until the determination is finished.
31.5 Manual control
Problem
Cause Remedy
Tapping on the fixed
key [ ] has no
effect.
The fixed key is disabled. In the Dialog options / Fixed keys dialog,
activate the Manual control option (see
chapter 7.2, page 24).
A button in the
manual control is
disabled.
The required hardware is
either not connected or
incorrectly connected.
1. Switch off the 900 Touch Control.
2. Connect the hardware correctly.
3. Switch the 900 Touch Control back on.

Table of Contents