22.2 Performing a sample series
■■■■■■■■■■■■■■■■■■■■■■
206
■■■■■■■■
900 Touch Control
Sequence of the determination
■ After a determination is started, a check is made as to
– whether all of the devices required are connected and ready for
operation.
– whether all titrants and sensors are present.
– whether the monitoring intervals for titrants, sensors and com-
mon variables are still valid.
– if a sample assignment table is activated (see chapter 15.1.2,
page 149):
whether the entered sample identification is defined in the sam-
ple assignment table. If this is the case, then the assigned
method will be loaded automatically and the actual determina-
tion will be started.
NOTICE
Manual titrations (MAT command):
Start of titration: Because of the technical implementation involved, a
few microliters are already dosed at the start of a manual titration. The
possibility of deviation from the actual endpoint volume is excluded,
however, because the already dosed volume is taken into account.
End of titration: A manual titration must always be ended with the
[Cancel command] button (NOT with the [ ] fixed key) or with the
[FILL] key on the Manual Dosing Controller. This is the only way to
ensure that subsequent method commands will be carried out.
22.2 Performing a sample series
You can use the following functions with sample series:
■ Statistics for multiple determinations
The statistics calculations are enabled in the method options (see chap-
ter 16.5, page 172).
■ Sample table
You can enter the sample data for an entire sample series in the sample
table (see chapter 21, page 197).