29.7 STAT titrations (STAT)
■■■■■■■■■■■■■■■■■■■■■■
368
■■■■■■■■
900 Touch Control
29.7.6 Control instrument
The control instrument the determination is carried out with is selected
under [Control device]. Control instruments are defined under Sys-
tem Device manager.
NOTICE
This button is only displayed when several control instruments have
been configured.
Control device
Selection of the control instrument from the list of devices. Only those
devices are displayed which are able to carry out the command. The con-
trol instruments do not have to be connected. This allows to transfer
methods more easily from one system to another.
Selection Selection of configured control instruments
29.7.7 Sensor
The parameters for the sensor are edited under [Sensor].
Measuring input
Selection of the measuring input the sensor is connected to. The selection
is not dependent on whether the control device has one or two measur-
ing interfaces.
Selection 1 | 2
Default value 1
Sensor
Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System Sensors. You can
also enter a sensor name which is not contained in the sensor list. When a
determination is started there is a check whether the sensor is contained
in the sensor list.
Selection Selection of configured sensors
Electrode check
on | off (Default value: off)
For the following electrodes, an electrode check can be carried out:
pH electrodes
Metal electrodes
Ion-selective electrodes