32.8 Stored buffer series for pH calibration
■■■■■■■■■■■■■■■■■■■■■■
506
■■■■■■■■
900 Touch Control
Temp.
(°C)
Beckmann
pH
4.00
pH
7.00
pH
10.01
50 4.06 6.97 9.83
55 4.08 6.98 -
60 4.09 6.98 -
65 4.11 6.99 -
70 4.12 6.99 -
75 4.14 7.00 -
80 4.16 7.00 -
85 4.18 7.01 -
90 4.19 7.02 -
95 4.21 7.03 -
NOTICE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.
However, they may be changed by the respective manufacturers.
32.8.8 Radiometer Analytical
Table 30
Radiometer Analytical buffer solutions
Temp.
(°C)
Radiometer (Analytical)
pH
1.679
pH
4.005
pH
7.000
pH
9.180
0 1.666 4.000 7.118 9.464
5 1.668 3.998 7.087 9.395
10 1.670 3.997 7.059 9.332
15 1.672 3.998 7.036 9.276
20 1.675 4.001 7.016 9.225
25 1.679 4.005 7.000 9.180
30 1.683 4.011 6.987 9.139
35 1.688 4.018 6.977 9.102
40 1.694 4.027 6.970 9.068

Table of Contents