EasyManua.ls Logo

Metrohm 900 - Editing Result Templates

Metrohm 900
553 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15.2 Custom result templates
■■■■■■■■■■■■■■■■■■■■■■
152
■■■■■■■■
900 Touch Control
15.2.1 Editing result templates
List of result templates: Result template New / Edit
Result name
The result name is the text which will be shown in the results display and
in the report.
Entry max. 24 characters
Default value R
Calculation formula
Shows the calculation formula. A special editor is opened for the defini-
tion (see chapter 29.13.3, page 427).
Entry max. 100 characters
Default value empty
Decimal places
Number of decimal places used to display the result.
Input range 0 to 5
Default value 2
Result unit
The result unit is displayed and saved together with the result.
Entry max. 10 characters
Selection % | mol/L | mmol/L | g/L | mg/L | mg/mL |
mg/100 g | ppm | g | mg | µg | mL | µL | mg/
piece | °C | µg/min | mL/min | µL/min
Default value %
[Note]
Entering a note on the calculation.
[Note for wizard]
Enter a note on the calculation. This note is displayed when loading the
result template.
[Result options]
Defining additional settings for the calculation.
"Edit result template / Result options" dialog
In the dialog Edit result template / Result options, settings for how to
process the calculated result are defined.

Table of Contents

Related product manuals