8 Titrants
900 Touch Control
■■■■■■■■
49
8.4.2 Tubing parameters
Titrant: Edit Dosing unit Tubing param.
You can enter the length and diameter of the connected tubings in the
dialog Dosing unit / Tubing parameters. The values which have
already been entered correspond to the dimensions of the supplied stan-
dard tubings. In addition, the port assignment can be modified.
NOTICE
These parameters are important for the correct execution of the Pre-
paring (PREP command) and Emptying (EMPTY command) functions,
because the volumes of the tubing connections are taken into account.
Dosing port 1
Port
Port to be used as dosing port 1 for the PREP and EMPTY (see figure 8,
page 51) functions.
Selection Port 1 | Port 2 | Port 3 | Port 4
Default value Port 1
Length
Length of the tubing.
Input range 0.0 to 999.9 cm
Default value 40.0 cm
The setting 0.0 means that this tubing will neither
be rinsed nor emptied.
Diameter
Diameter of the tubing.
Input range 0.0 to 9.9 mm
Default value 2.0 mm
Dosing port 2
Port
Port to be used as dosing port 2 for the PREP and EMPTY functions (see
figure 8, page 51).
Selection Port 1 | Port 2 | Port 3 | Port 4
Default value Port 3