EasyManua.ls Logo

Metrohm 900 - Coulometric Karl Fischer Titration

Metrohm 900
553 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
■■■■■■■■■■■■■■■■■■■■■■
31 Troubleshooting
900 Touch Control
■■■■■■■■
489
31.9 Coulometric Karl Fischer titration
Problem Cause Remedy
The drift is very high
during conditioning.
The titration cell is leaking. Check the septum and replace it if neces-
sary.
Replace the molecular sieve.
Check the ground-joint sleeves for frayed
edges. If necessary, trim the edges neatly
or replace the ground-joint sleeves.
Deposits containing water
can be found in the titra-
tion cell.
Shake the titration cell.
The reagent is exhausted
or contaminated.
Replace the reagent.
The catholyte is old or
moist.
The catholyte must be replaced at least
once per week.
Dry the catholyte with KF one-component
reagent.
The diaphragm of the gen-
erator electrode is contam-
inated.
Clean the diaphragm (in accordance with
directions in the leaflet for the generator
electrode).
A side reaction is taking
place.
Consider using a KF oven.
See technical literature.
The molecular sieve on the
KF oven has reached its
capacity limit.
Replace the molecular sieve.
The gas flow from the KF
oven into the titration cell
is too high.
Reduce the gas flow (if you are working
with the oven method with sample vials,
set the gas flow between 40 and 60
mL/min).
The drift becomes
greater after each
titration.
The sample releases water
very slowly.
Consider using a KF oven.
See technical literature.
The drift is fluctuat-
ing.
The sample solution is
poorly stirred.
Set the stirring rate in such a way that the
sample solution will become very well
mixed.

Table of Contents

Related product manuals