■■■■■■■■■■■■■■■■■■■■■■
31 Troubleshooting
900 Touch Control
■■■■■■■■
489
31.9 Coulometric Karl Fischer titration
Problem Cause Remedy
The drift is very high
during conditioning.
The titration cell is leaking. ■ Check the septum and replace it if neces-
sary.
■ Replace the molecular sieve.
■ Check the ground-joint sleeves for frayed
edges. If necessary, trim the edges neatly
or replace the ground-joint sleeves.
Deposits containing water
can be found in the titra-
tion cell.
■ Shake the titration cell.
The reagent is exhausted
or contaminated.
■ Replace the reagent.
The catholyte is old or
moist.
■ The catholyte must be replaced at least
once per week.
■ Dry the catholyte with KF one-component
reagent.
The diaphragm of the gen-
erator electrode is contam-
inated.
■ Clean the diaphragm (in accordance with
directions in the leaflet for the generator
electrode).
A side reaction is taking
place.
■ Consider using a KF oven.
■ See technical literature.
The molecular sieve on the
KF oven has reached its
capacity limit.
■ Replace the molecular sieve.
The gas flow from the KF
oven into the titration cell
is too high.
■ Reduce the gas flow (if you are working
with the oven method with sample vials,
set the gas flow between 40 and 60
mL/min).
The drift becomes
greater after each
titration.
The sample releases water
very slowly.
■ Consider using a KF oven.
■ See technical literature.
The drift is fluctuat-
ing.
The sample solution is
poorly stirred.
■ Set the stirring rate in such a way that the
sample solution will become very well
mixed.