Metrohm 900 - Monitoring the Calibration Interval (Only for Ph and Ion-Selective Electrodes and Conductivity Measuring Cells)

Metrohm 900
553 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
■■■■■■■■■■■■■■■■■■■■■■
10 Sensors
900 Touch Control
■■■■■■■■
77
If this parameter is activated, then the cell constant will be monitored.
Lower limit
Input range 0.001 to 500.000 /cm
Default value 0.850 /cm
Upper limit
Input range 0.001 to 500.000 /cm
Default value 1.150 /cm
10.6 Monitoring the calibration interval (only for pH and
ion-selective electrodes and conductivity measuring
cells)
Sensor: Edit Calibration interval
In the dialog Edit sensor / Calibration interval, you can define the time
interval after which the sensor must be recalibrated.
Calibration date
Date of the last calibration.
Monitoring
on | off (Default value: off)
If this parameter is activated, then the validity of the calibration will be
monitored.
Calibration interval
If you define a time interval for the validity of the calibration, then the
date in Next calibration will be tracked automatically.
Input range 1 to 999 days
Default value 7 days
Next calibration
If you define a date for the next calibration, then the Calibration inter-
val will be tracked automatically.
Format: YYYY:MM:DD
Action
Selection of the action which is carried out when the time interval has
expired.