29 Parameters
900 Touch Control
■■■■■■■■
353
Electrode check
on | off (Default value: off)
For polarizable electrodes and the generator electrode, an electrode check
can be carried out. The following will be tested:
■ Polarizable electrodes
– Is the electrode connected?
– Is there a short circuit?
The electrode check is carried out when this command is started.
■ Generator electrode:
– Is the electrode connected?
– Is the generator system defective?
– Is the resistance of the sample solution very high?
The electrode check is carried out continuously during the execution of
this command.
Temp. measurement
Type of temperature measurement.
Selection continuous | automatic | off
Default value automatic
continuous
A temperature sensor must be connected. The temperature is mea-
sured continuously.
automatic
If a temperature sensor is connected then the temperature will be mea-
sured continuously. Otherwise, the temperature entered manually will
be used (see dialog of the titration and measuring parameters).
off
The temperature will not be measured. The temperature entered man-
ually is used (see dialog of the titration and measuring parameters).
29.6.8 Stirrer
The parameters for the stirrer are edited under [Stirrer].
Stirrer
Selection of the MSB connector the stirrer is connected to. Every MSB
connector will always be displayed.
Selection 1 | 2 | 3 | 4 | off
Default value 1
off
No stirrer will be used.