29.8 Manual titrations (MAT)
■■■■■■■■■■■■■■■■■■■■■■
374
■■■■■■■■
900 Touch Control
29.8.1 Dosing parameters
Under [Dosing parameters], the parameters influencing the run of the
manual titration are defined.
Dosing ramp
The dosing ramp is a delay in the dosing rate at the time of the start of a
dosing step. It is advantageous primarily for manual titrations with indica-
tor, when small volume steps are to be dosed before the equivalence
point. The selectable delay time determines after which amount of time
the defined dosing rate is to be reached.
Input range 0 to 10 s
Default value 2 s
Dosing rate
The rate at which dosing takes place.
Input range 0.01 to 166.00 mL/min
Selection maximum
Default value maximum
maximum
The maximum dosing rate depends on the cylinder volume (see chap-
ter 32, page 492).
Time interval MP
Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.
Input range 0.1 to 999999.0 s
Default value 5.0 s
Filling rate
Rate at which the dosing cylinder is filled.
Input range 0.01 to 166.00 mL/min
Selection maximum
Default value maximum
maximum
The maximum filling rate depends on the cylinder volume (see chapter
32, page 492).
Temperature
Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement = automatic or continuous is defined (see sensor dialog