29.7 STAT titrations (STAT)
■■■■■■■■■■■■■■■■■■■■■■
360
■■■■■■■■
900 Touch Control
Input range 0.01 to 166.00 mL/min
Selection off
Default value off
NOTICE
If several of the parameters Start time, Start meas. value or Start
rate have been defined, then all of them must be fulfilled for measured
values to be entered in the measuring point list.
29.7.4 Stop conditions
The conditions for canceling the titration are defined under [Stop condi-
tions].
Stop volume
The titration is stopped when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.
Input range 0.00000 to 9999.99 mL
Default value 100,000 mL
Selection off
Stop time
The titration is stopped when the specified time has elapsed.
Input range 0 to 999999 s
Selection off
Default value off
Since
This parameter can only be edited when a stop time has been defined.
This parameter defines the start of the stop time.
Selection Start | Control point | Last dosing
Start
The stop time starts after the start volume has been dosed and the fol-
lowing pause time has elapsed.
Control point
The stop time starts as soon as the control point has been reached for
the first time.
Last dosing
The stop time starts after the last increment has been dosed, i. e. the
stop time is reset to zero again after each dosing step.