29 Parameters
900 Touch Control
■■■■■■■■
359
Temperature
Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement = automatic or continuous is defined (see sensor dialog
of the command). This value is used for temperature correction in pH
measurements.
Input range –20.0 to 150.0 °C
Default value 25.0 °C
Time interval MP
Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.
Input range 0.1 to 999999.0 s
Default value 2.0 s
Start time
The measured values are not entered in the measuring point list until this
time has elapsed since the start of the titration.
Input range 0 to 999999 s
Default value 0 s
Start meas. value
The measured values are not entered in the measuring point list until this
start measured value has been reached.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
instrument used.
Measuring mode pH:
Input range –20.000 to 20.000
Selection off
Default value off
Measuring mode U:
Input range
–2000.0 to 2000.0 mV
Selection off
Default value off
Start rate
The measured values are not entered in the measuring point list until the
dosing rate has fallen below this value. Checking this parameter is not
started until 10 s after the start of the command.