29.7 STAT titrations (STAT)
■■■■■■■■■■■■■■■■■■■■■■
362
■■■■■■■■
900 Touch Control
can be found in the "Technical specifications" chapter for the control
instrument used.
Measuring mode pH:
Input range –20.000 to 20.000
Default value –20.000
Measuring mode U:
Input range –2000.0 to 2000.0 mV
Default value –2000.0 mV
Dosing rate:
Input range
0.00 to 166.00 mL/min
Default value 0.00 mL/min
Temperature:
Input range
–20.0 to 150.0 °C
Default value –20.0 °C
Lower hysteresis
The lower hysteresis represents the tolerance range for the lower limit (see
figure 18, page 365). When Action = Wait the dosing does not continue
until the monitored quantity has once again exceeded the lower limit incl.
the value entered here.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
instrument used.
Measuring mode pH:
Input range
0.000 to 20.000
Default value 0.020
Measuring mode U:
Input range
0.0 to 2000.0 mV
Default value 2.0 mV
Dosing rate:
Input range
0.00 to 166.00 mL/min
Default value 0.20 mL/min
Temperature:
Input range
0.0 to 150.0 °C
Default value 0.2 °C