29.10 Calibration of sensors (CAL)
■■■■■■■■■■■■■■■■■■■■■■
396
■■■■■■■■
900 Touch Control
Remote
For automatic calibrations with Sample Processors, which are con-
nected via the Remote Box.
USB
For automatic calibrations with a Sample Processor that is connected
via USB(see "Definition of the calibration positions", page 498).
Standards
The concentrations of the standards as well as their units are defined
under [Standards].
Unit
Unit of the standard solutions used. For concentration measurements
without standard addition, this unit will be used.
Selection mol/L | mmol/L | ppm | % | g/L | mg/L | µg/L
Default value ppm
Concentration standard 1
Concentration of the standard solution used.
Input range 00.001 to 999999.999
Default value 1,000
Concentration standard 2
Concentration of the standard solution used.
Input range 00.001 to 999999.999
Selection off
Default value off
Concentration standard 3
See Concentration standard 2.
Concentration standard 4
See Concentration standard 2.
Concentration standard 5
See Concentration standard 2.
Stirrer control
The parameters for the stirrer control during the calibration are edited
under [Stirrer control]. Ensure that a stirrer has been selected in the Edit
command / Stirrer dialog.