Defining Parameter Files ASAP 2460 Operator’s Manual
2-20 246-42800-01 - Aug 2013
3. Click OK when prompted to create a new file.
4. To overwrite run conditions with parameters from another Analysis Conditions file, click the
Description dropdown arrow and select a file from the list or click Browse, then locate and select
the .ANC file containing the new parameters and click Open.
5. To overwrite adsorptive properties with parameters from another Adsorptive Properties file, click
the Adsorptive dropdown arrow and select a file from the list or click Browse, then locate and
select a .ADP file containing the new parameters and click Open. Refer to Adsorptive
Properties, page 2-21 and Analysis Conditions Files, page 3-15.
6. Use the Insert Range button to enter starting and ending relative pressure points.
7. Select the Absolute pressure dosing checkbox to specify pressure targets in mmHg, mbar, or kPa
instead of relative pressure. This option is typically selected when using adsorptives at analysis
conditions above the critical point of the gas; for example, H
2
adsorption on carbon at liquid
nitrogen temperature.
8. Click the following buttons to specify:
9. Click Save, then click Close.
Button Use to Specify...
Preparation evacuation rate/time/level, leak test and time values, elevator
prompts
Free space how the free space is to be measured
pº and T
how the saturation pressure (
p
o
) is to be measured and the
analysis bath temperature
Equilibration equilibration intervals
Backfill backfill options at start of analysis and after analysis