Apendix E ASAP 2460 Operator’s Manual
E-2 246-42800-01 - Aug 2013
Purge Method
This is the preferred method for removing trapped air.
1. Go to Unit > Enable Manual Control (if the instrument schematic is not displayed, select Show
Instrument Schematic).
2. Close all valves by right clicking on each valve and selecting Close.
3. Open the regulator Shut-off valve.
If multiple instruments are installed, make sure to choose the correct Unit menu.