Appendix B ASAP 2460 Operator’s Manual
B-34 246-42800-01 - Dec 2013
6200 Series
6204- This sample has an invalid status and cannot be used for degassing.
Cause: You tried to degas a sample that has a status other than No Analysis.
Action: Select a sample with a status of No Analysis.
6206- Error dosing.
Cause: The degas system failed to dose the backfill gas to a pressure of 850
mmHg.
Action: The gas regulator may be set too
low or turned off.
6212- Calibrating. This action must be completed before the application exits.
Cause: The unit is calibrating.
Action: Do not exit the application until the
calibration is complete.
6213- Error overheating on port (number). Current = (temperature), Target = (temper-
ature), Limit = (temperature).
Cause: A heating mantle and/or its thermocouple are not plugged into the cor-
rect port.
Action: Verify heating mantle connections.
If the problem persists, contact
your service representative.
6217- Error thermocouple unplugged Current = (temperature), Target = (temperature).
Cause: The thermocouple is not plugged into the correct degas port.
Action: Check the heating mantle connections for the
specified degas port.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ASAP 2460 and is the answer not in the manual?

Micromeritics ASAP 2460 Specifications

General IconGeneral
BrandMicromeritics
ModelASAP 2460
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals