Related Topics
See "QuickStart Degas Conditions" on page 3 - 2
See "Start Smart VacPrep Degas" on page 3 - 5
See "Set the Voltage for the Smart VacPrep" on the next page
SPECIFICATIONS FOR THE SMART VACPREP
The Smart VacPrep degasser has been designed and tested to meet the following specifications:
Computer
Windows 7 Professional or higher operating system is recommended for the best user
experience. If the computer is to be connected to a network, a second Ethernet port on the
computer must be used for that purpose.
Specification Description
Backfill Options Automatically backfill or wait for user command
Capacity 6 ports
Degas Method Heat and vacuum
Physical Dimensions Height: 69 cm (27 in.)
Width: 53 cm (20.75 in.)
Depth: 44 cm (17 in.)
Weight: 32 kg (70 lbs)
Evacuation Period Determined by selected elapsed time or achievement of selected low pres-
sure target
Evacuation Rate Selectable evacuation rate from 1.0 to 50.0 mmHg/s
Heating of Samples Heating mantles
Heating Ramp Rate Selectable from 0.1
o
C per minute to 10
o
C per minute
Soak Time Zero to 999 minutes in one minute increments (Zero = Manual Cancel).
Temperature Accuracy Deviation less than ±10
o
C at the sensing thermocouple embedded in the
heating mantle
Temperature Control One ramp during evacuation phase, five additional ramps during sub-
sequent phases
Temperature Range 40 to 450
o
C in one degree increments
Vacuum Control Selectable target pressure controls switchover from restricted to unres-
tricted evacuation.
Specifications for the Smart VacPrep
Specifications for the Smart VacPrep
Smart VacPrep Operator Manual 1 - 3