ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-17
4000 Series
4002- Thermal Transpiration correction had no effect.
Cause: The Thermal transpiration correction option was selected on the Report
Options dialog box. However, the correction did not change any pres-
sure by more than one percent.
Action: De-select this option to disable this me
ssage. This correction is only
meaningful for very low pressures.
4003- Error Converting Pressures.
4004- Error Computing Quan
tity Adsorbed.
Cause: An internal processing and/or hardware error occurred during report
generation.
Action: Contact your service representative if you continue to receive this error
me
ssage.
4005- Pressures were not smoothed. Not enough pressures below 0.10 P/Po.
Cause: The Smooth pressures below 0.10 P/Po option was selected on the
Report Options dialog. There must be at least 10 pressures within this
range for smoothing to occur.
Action: Deselect this option to disable this message.
4011- Analysis gas in sample file does not match analysis gas in unit.
Cause: You attempted to start an analysis using a sample information file in
which the analysis gas specified does not match the analysis gas
entered in the unit configuration.
Action: If necessary, attach the appropriate gas bottle, then enter the gas in the
Un
it Configuration dialog box.