EasyManua.ls Logo

Micromeritics ASAP 2460 - Page 298

Default Icon
468 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix B ASAP 2460 Operator’s Manual
B-18 246-42800-01 - Dec 2013
4012- Psat gas in sample file does not match Psat gas in unit.
Cause: You attempted to start an analysis using a sample information file in
which the Psat gas does not match the Psat gas entered in the unit con-
figuration.
Action: If necessary, attach the appropriate gas bottle, then enter the gas in the
Un
it Configuration dialog box.
4014- File (name) is not a valid file for conversion.
Cause: The file selected for conversion is not a MICMOS 2400 or 2405 file.
Action: Select only files that have been created by MICMOS 2400 or 2405
p
rograms.
4015- Error creating export file for sample (filename).
Cause: A file error occurred during creation of an export output file.
Action: The output file name may be invalid. Ensure
that the target directory
exists. Ensure that the target diskette or CD is not full or write pro-
tected. The target disk drive may be damaged or inoperative. Verify
tha
t other files may be created on the same drive. Contact your service
representative if you continue to receive this error message.
4016- Sample (file name) has no data for export.
Cause: The file selected for export has a status of No Analysis. No export file
will be created.
Action: Select a file which contains
analysis data.

Table of Contents

Related product manuals