Page #298 background image
Appendix B ASAP 2460 Operator’s Manual
B-18 246-42800-01 - Dec 2013
4012- Psat gas in sample file does not match Psat gas in unit.
Cause: You attempted to start an analysis using a sample information file in
which the Psat gas does not match the Psat gas entered in the unit con-
figuration.
Action: If necessary, attach the appropriate gas bottle, then enter the gas in the
Un
it Configuration dialog box.
4014- File (name) is not a valid file for conversion.
Cause: The file selected for conversion is not a MICMOS 2400 or 2405 file.
Action: Select only files that have been created by MICMOS 2400 or 2405
p
rograms.
4015- Error creating export file for sample (filename).
Cause: A file error occurred during creation of an export output file.
Action: The output file name may be invalid. Ensure
that the target directory
exists. Ensure that the target diskette or CD is not full or write pro-
tected. The target disk drive may be damaged or inoperative. Verify
tha
t other files may be created on the same drive. Contact your service
representative if you continue to receive this error message.
4016- Sample (file name) has no data for export.
Cause: The file selected for export has a status of No Analysis. No export file
will be created.
Action: Select a file which contains
analysis data.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ASAP 2460 and is the answer not in the manual?

Micromeritics ASAP 2460 Specifications

General IconGeneral
BrandMicromeritics
ModelASAP 2460
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals