Report Options Files ASAP 2460 Operator’s Manual
3-72 246-42800-01 - Aug 2013
Options Report
The Options report lists the conditions used to perform the analysis. It contains analysis information
including:
Degas conditions
Adsorptive properties
Analysis conditions
Free space
Saturation pressure ( p
o
) and temperature
Analysis method
Sample Log Report
In the Selected Reports list box, select Sample Log. This report provides the following information:
Manual control operations performed during analysis
Information entered using Add Log Entry on the sample file editor
Warnings and/or errors which occurred during analysis
Validation Report
In the Selected Reports list box, highlight Validation, then click Edit. This report allows data to be
examined by the analysis program to determine if the results are within typical ranges. If the data for
any reports selected for validation are determined to be out of range, a warning displays, and sugges-
tions are given for corrective action. This information is detailed in the report and plotted on the graph
as a unique plot symbol.