ASAP 2460 Operator’s Manual Start Micropore Analysis
246-42800-01 - Aug 2013 4-13
Start Micropore Analysis
When you select this option, the Micropore Analysis dialog is displayed. This dialog enables you to
start up to six analyses at the same time. Use the following guidelines for performing micropore
analyses:
• All ports must be idle in order to start an analysis.
• All analyses must use the same analysis gas.
• All analyses that measure Psat must use the same Psat gas (which may be different from the analysis
gas).
• The sample dosing method should be Normal.
• From one to six analyses can be started at the same time.
• Samples cannot be removed from or added to ports until the full set of analyses has completed.
The fields in this window are the same as in the Krypton Analysis window. Refer to Start Krypton
Analysis, page 4-8.