Routine Maintenance ASAP 2460 Operator’s Manual
7-24 246-42800-01 - Aug 2013
Specifying Gas Ports
When you change the gas attached to a port on the analyzer, you must enter the change in the analysis
program as follows.
1. Select Unit [n] > Unit Configuration from the main menu. The Unit Configuration dialog is
displayed.
2. Click on the down-arrow at the port to which you attached a gas line and choose the appropriate
gas.
3. Click
OK to close the dialog, and then click OK again to close the Unit Configuration dialog.