Start Krypton Analysis ASAP 2460 Operator’s Manual
4-8 246-42800-01 - Aug 2013
Start Krypton Analysis
When you select this option from the Unit menu, the Krypton Analysis dialog is displayed.
Use the following guidelines for performing krypton analyses:
All ports must be idle in order to start an analysis.
All analyses must use krypton gas.
From one to five analyses can be started at the same time using ports 1 through 5. Port 6 is reserved
for dosing.
A blank tube must be installed in the last port (2, 4, or 6 depending on your system configuration).
The sample file used for an analysis must specify the Dosing Method: From last port in the
Adsorptive Properties.
Samples cannot be removed from or added to ports until the full set of analyses has
completed.