Start Krypton Analysis ASAP 2460 Operator’s Manual
4-8 246-42800-01 - Aug 2013
Start Krypton Analysis
When you select this option from the Unit menu, the Krypton Analysis dialog is displayed.
Use the following guidelines for performing krypton analyses:
• All ports must be idle in order to start an analysis.
• All analyses must use krypton gas.
• From one to five analyses can be started at the same time using ports 1 through 5. Port 6 is reserved
for dosing.
• A blank tube must be installed in the last port (2, 4, or 6 depending on your system configuration).
• The sample file used for an analysis must specify the Dosing Method: From last port in the
Adsorptive Properties.
• Samples cannot be removed from or added to ports until the full set of analyses has 
completed.