Preventive Maintenance ASAP 2460 Operator’s Manual
7-10 246-42800-01 - Aug 2013
Changing Fluid
1. Unplug the vacuum pump from the power source.
2. Loosen the wing nut on the clamp at the top of the oil vapor trap. Swing the clamp open and
remove the trap from the hose.
Always drain the vacuum pump while the pump is warm and disconnected from
the power source.
Disconnect hose
Oil-vapor trap
Clamp wing-nut