EasyManuals Logo

Micromeritics ASAP 2460 User Manual

Default Icon
468 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #328 background imageLoading...
Page #328 background image
Appendix C ASAP 2460 Operator’s Manual
C-4 246-42800-01 - Aug 2013
Free Space
Measured
Measured free-space volumes are calculated using the following equations:
where:
P
1
= system manifold pressure before dosing helium onto sample
P
2
= system manifold pressure after dosing helium onto sample
P
3
= sample pressure after raising dewar and equilibrating with helium
T
amb
= approximate room temperature (298 K)
T
bath
= analysis bath temperature (K)
T
man
= system manifold temperature before dosing helium onto sample (K)
T
STD
= standard temperature (273.15 K)
V
bath
= portion of cold free space at analysis bath temperature
V
fc
= volume of free space, cold (cm
3
at standard temperature)
V
fw
= volume of free space, warm (cm
3
at standard temperature)
V
man
= manifold volume (cm
3
)
V
fw
V
man
T
man
------------
P
1
P
2
------
1–


T
STD
=
V
fc
V
fw
P
2
P
3
------


=
V
bath
V
fc
V
fw
–
1
T
bath
T
amb
------------
–
----------------------
=

Table of Contents