4 Smart VacPrep Calibration
SMART VACPREP SERVO VALVE CALIBRATION
Smart VacPrep >Unit [n] > Calibration >Servo Valve
Use to calibrate the servo valve to the manifold pressure transducer. The servo valve should always be
recalibrated after a pressure calibration has been performed. The pressure transducer should be
calibrated before starting this calibration procedure.
1. Go to Smart VacPrep > Unit [n] > Calibration > Servo Valve.
2.
Click Start. The window closes when the calibration is complete. Click Cancel to stop the cal-
ibration process.
Field or Button Description
Cancel Discards any changes or cancels the current process.
Start Starts the operation.
Servo Valve Fields and Buttons Table
Related Topics
See "Smart VacPrep Calibration" on the previous page
See "Smart VacPrep Pressure Offset" on the previous page
4 - 2 Smart VacPrep Operator Manual