ASAP 2460 Operator’s Manual Defining Parameter Files
246-42800-01 - Aug 2013 2-19
4. To overwrite degas conditions with parameters from another Degas Conditions file, click the
Degas Conditions dropdown arrow and select a file from the list or click Browse, then locate and
select the .DEG file containing the new parameters and click Open.
5. Use the Insert button to enter up to five stages of degassing (soak temperature, ramp rate, and
soak time). When using the SmartPrep degasser, the maximum temperature that can be entered is
450 °C.
6. Click Save, then click Close.
Analysis Conditions
Analysis conditions specify the data used to guide an analysis.
1. Go to File > Open. Select
the Analysis Conditions library folder and enter a file name in the File
name field.
2. Click Open.