Appendix B ASAP 2460 Operator’s Manual
B-10 246-42800-01 - Dec 2013
2498- The requested change to the Sample’s status is invalid at this time.
Cause: A request to change the file’s status (for example, from automatically
collected to manually entered) could not be done.
Action: Contact your service representative if yo
u continue to receive this error
message. Record the name of the sample file in which the problem
occurred.
2499- Sequence number must contain at least 3 digits.
Cause: You tried to enter a sequence number that did not contain at least three
digits.
Action: Enter a sequence number that contains at least three digits.
2500 Series
2500- All sample file names that can be created using the sequence number pattern
already exist. You may want to modify the next sequence number.
Cause: No more sample information files can be created using the currently
entered file name sequence number.
Action: Select Op
tions > Sample Defaults from the Main Menu and enter a
new sequence number.
2501- System resources have reached a dangerously low level. Please close some
windows to avoid the loss of data.
Cause: A large number of windows are open and consuming the system
resources available to all applications.
Action: Close one or more windows on the screen. Contact your service rep-
resentative if you continue to receive this error message.
2505- Error Logger cannot be initialized! Error code (number). Program will exit.
Cause: An internal processing and/or hardware error has occurred.