ASAP 2460 Operator’s Manual Start Krypton Analysis
246-42800-01 - Aug 2013 4-9
Field or Button Description
View dropdown list Refer to Common Fields and Buttons - Unit Menu Options,
page 4-2
Browse button Refer to Common Fields and Buttons - Unit Menu Options,
page 4-2
Ports 1 through 6 Allows you to select the ports to be used for the analysis. The last
port (2, 4, or 6, based on your system configuration) must contain
a blank sample tube to be used for krypton dosing.
Sample text field Displays the sample’s file name.
Mass group box
(continued)
Calculate - enables the Empty tube and Sample + tube
fields. Enter the values necessary to calculate the sample mass.
Equation used to calculate sample mass:
Mass
sample
= Mass
sample+tube
– Mass
tube
Density - value is used for the Calculated free space method
only. Use 0.000 for a blank analysis.
p
o
text field Enabled if you have chosen to enter the p
o
measurement. Enter (or
confirm) the saturation pressure of the adsorptive, or edit the
current one if desired.
Bath temperature text field
Enabled if you have chosen to enter the bath temperature ( p
o
and
T option on the Analysis Conditions dialog). Enter (or confirm)
the temperature for the analysis bath.
Report After Analysis button Refer to Common Fields and Buttons - Unit Menu Options,
page 4-2.
Start button Begins the analysis.