Appendix B ASAP 2460 Operator’s Manual
B-16 246-42800-01 - Dec 2013
2563- Cannot write. File or directory (file/directory name) read only.
Cause: The referenced file or directory is read-only. You cannot write to this
file.
Action: Have the administrator of the compute
r set up proper read/write access
to the directories and files.
2572- File (filename) is already selected for the analysis.
Cause: The file selected for HOA reporting cannot produce a valid report.
Action: Choose a different sample file.
2585- The following libraries are missing: <library-name>, <library-name>, ....
2586- Sample file [name] has no pressure table entries.
2587- Sample file [name] has no dose incremental table entries.
2599- The selected file has an extension that is not supported by this operation.
Cause: This message is triggered on application start up if any of the library
files used by an application, does not exist on disk.
Action: Add the library into the Libraries folder.
Cause: An attempt was made to start an analysis with an empty target
pressure table.
Action: Add target pressures to the pressure table in the analysis conditions.
Cause: An attempt was made to start an analysis with an empty dose
increment table.
Action: Add dose increment entries to the dose increment table in the analysis
conditions.
Cause: The selected file does not havea a supprted file extension.
Action: Open the adsorptive properties file. Open the FPI file selector and
select another file with a supported file extension.